T1 - Use of a blood glucose meter for radiochromic film analysis in blood irradiation
AU - Cheung, Tsang
AU - Butson, Martin J.
AU - Yu, Peter K. N.
PY - 2002/10/7
Y1 - 2002/10/7
N2 - The use of a diabetic blood glucose meter for radiochromic film dosimetry in blood irradiation using x-ray beams on a medical linear accelerator has been investigated. The glucose meter provides optical density analysis in the visible and infrared region using a reflectance measurement technique. By comparing the 'blood sugar' level output with standard calibration gafchromic films a calibration curve is produced for quantitative analysis. Results show that a reproducible dose to meter output curve can be fitted using a second order polynomial function and that blood irradiation doses in vitro were measured to within 7.9% mean error (as compared to ionization chamber results) using the blood glucose meter. This level of accuracy falls below that measured with a standard densitometer (4.3%); however, results show that the blood glucose meter, which would be available in any haematology department, produces an adequate measure of gafchromic film optical density for blood irradiation dosimetry.
AB - The use of a diabetic blood glucose meter for radiochromic film dosimetry in blood irradiation using x-ray beams on a medical linear accelerator has been investigated. The glucose meter provides optical density analysis in the visible and infrared region using a reflectance measurement technique. By comparing the 'blood sugar' level output with standard calibration gafchromic films a calibration curve is produced for quantitative analysis. Results show that a reproducible dose to meter output curve can be fitted using a second order polynomial function and that blood irradiation doses in vitro were measured to within 7.9% mean error (as compared to ionization chamber results) using the blood glucose meter. This level of accuracy falls below that measured with a standard densitometer (4.3%); however, results show that the blood glucose meter, which would be available in any haematology department, produces an adequate measure of gafchromic film optical density for blood irradiation dosimetry.
UR - http://www.scopus.com/inward/record.url?scp=0037037756&partnerID=8YFLogxK
UR - https://www.scopus.com/record/pubmetrics.uri?eid=2-s2.0-0037037756&origin=recordpage
U2 - 10.1088/0031-9155/47/19/401
DO - 10.1088/0031-9155/47/19/401
M3 - RGC 21 - Publication in refereed journal
C2 - 12408483
SN - 0031-9155
VL - 47
JO - Physics in Medicine and Biology
JF - Physics in Medicine and Biology
IS - 19
ER -