Abstract
Strong fat signal in regions where a large susceptibility difference exists, for instance at the interface between air and tissue near the maxillary sinus, may not be eliminated by currently available fat suppression techniques without sacrificing the overall quality of the images. In this article, we show that this fat signal, which appears as a susceptibility artifact, can be significantly reduced by using an optimized presaturation pulse with sharp edges and a broad bandwidth, while causing minimal disturbance of the water signal. Several optimized presaturation pulses can be reproduced by the Fourier coefficients provided in the Appendix.
| Original language | English |
|---|---|
| Pages (from-to) | 582-587 |
| Journal | Magnetic Resonance in Medicine |
| Volume | 33 |
| Issue number | 4 |
| DOIs | |
| Publication status | Published - 1995 |
| Externally published | Yes |
Research Keywords
- bandwidth of optimized saturation pulse
- susceptibility artifact
- width of the transition edge of response profile
Fingerprint
Dive into the research topics of 'Susceptibility artifact reduction in fat suppression'. Together they form a unique fingerprint.Cite this