Strong fat signal in regions where a large susceptibility difference exists, for instance at the interface between air and tissue near the maxillary sinus, may not be eliminated by currently available fat suppression techniques without sacrificing the overall quality of the images. In this article, we show that this fat signal, which appears as a susceptibility artifact, can be significantly reduced by using an optimized presaturation pulse with sharp edges and a broad bandwidth, while causing minimal disturbance of the water signal. Several optimized presaturation pulses can be reproduced by the Fourier coefficients provided in the Appendix.
Original languageEnglish
Pages (from-to)582-587
JournalMagnetic Resonance in Medicine
Volume33
Issue number4
DOIs
Publication statusPublished - 1995
Externally publishedYes

Research Keywords

  • bandwidth of optimized saturation pulse
  • susceptibility artifact
  • width of the transition edge of response profile