Rotating and static sources for gamma knife radiosurgery systems : Monte Carlo studies
Research output: Journal Publications and Reviews › RGC 21 - Publication in refereed journal › peer-review
Author(s)
Detail(s)
Original language | English |
---|---|
Pages (from-to) | 2500-2505 |
Journal / Publication | Medical Physics |
Volume | 33 |
Issue number | 7 |
Publication status | Published - Jul 2006 |
Link(s)
Abstract
Rotating gamma systems (RGSs), GammaART-6000™, and its Chinese equivalents, such as OUR and MASEP, etc., are new radiosurgery systems that use rotating 60Co sources instead of the 201 static sources (Leksell gamma knife, LGK). The rotating sources of RGSs simulate an infinite number of beams and promote extremely high target to surface dose ratios. However, the results of Monte Carlo in this study shows that RGS variants (modeled as having the same latitude angles, source to focus distance, and the distance from the source to the end of the collimator as the LGK) have smaller beam profile penumbra in the z direction, while LGK has smaller penumbra in the x and y directions. The differences are more significant in using larger collimators. © 2006 American Association of Physicists in Medicine.
Research Area(s)
- Gamma knife, Monte Carlo simulation, Stereotactic radiosurgery
Citation Format(s)
Rotating and static sources for gamma knife radiosurgery systems: Monte Carlo studies. / Cheung, J. Y C; Yu, K. N.
In: Medical Physics, Vol. 33, No. 7, 07.2006, p. 2500-2505.
In: Medical Physics, Vol. 33, No. 7, 07.2006, p. 2500-2505.