Abstract
Gafchromic XRCT, radiochromic film is assessed over a broad energy range, from kilovoltage to megavoltage x rays for variations in reflected optical density to dose response. A large energy dependence was found with reflected optical density output for the same delivered dose varying from 7.8±0.35 at 25.5 keV (50 kVp) peaking at 12.1±0.5 at 54 keV (125 kVp) to 0.975±0.03 at 2300 keV (10 MV) when normalized to 1 at 1400 keV (6 MV) energy. The response is constant (within 3%) in the 36-69 keV equivalent photon energy range, which corresponds to x-ray tube generating potentials of approximately 100-150 kVp. This matches well with beam qualities for diagnostic computed topography applications. © 2006 American Association of Physicists in Medicine.
| Original language | English |
|---|---|
| Pages (from-to) | 2923-2925 |
| Journal | Medical Physics |
| Volume | 33 |
| Issue number | 8 |
| DOIs | |
| Publication status | Published - 2006 |