Measurement of energy dependence for XRCT radiochromic film

Research output: Journal Publications and ReviewsRGC 21 - Publication in refereed journalpeer-review

29 Scopus Citations
View graph of relations

Author(s)

Detail(s)

Original languageEnglish
Pages (from-to)2923-2925
Journal / PublicationMedical Physics
Volume33
Issue number8
Publication statusPublished - 2006

Abstract

Gafchromic XRCT, radiochromic film is assessed over a broad energy range, from kilovoltage to megavoltage x rays for variations in reflected optical density to dose response. A large energy dependence was found with reflected optical density output for the same delivered dose varying from 7.8±0.35 at 25.5 keV (50 kVp) peaking at 12.1±0.5 at 54 keV (125 kVp) to 0.975±0.03 at 2300 keV (10 MV) when normalized to 1 at 1400 keV (6 MV) energy. The response is constant (within 3%) in the 36-69 keV equivalent photon energy range, which corresponds to x-ray tube generating potentials of approximately 100-150 kVp. This matches well with beam qualities for diagnostic computed topography applications. © 2006 American Association of Physicists in Medicine.

Research Area(s)

  • Dose response, Energy dependence, Gafchromic XRCT, Radiation dosimetry, Radiochromic film

Citation Format(s)

Measurement of energy dependence for XRCT radiochromic film. / Butson, Martin J.; Cheung, Tsang; Yu, Peter K. N.
In: Medical Physics, Vol. 33, No. 8, 2006, p. 2923-2925.

Research output: Journal Publications and ReviewsRGC 21 - Publication in refereed journalpeer-review