    Titanium oxide film with a graded interface to NiTi matrix was synthesized in situ on NiTi shape memory alloy(SMA) by oxidation in H2O2 solution. In vitro studies including contact angle measurement, hemolysis, MTT cytotoxicity and cell morphology tests were employed to investigate the biocompatibility of the H2O2-oxidized NiTi SMAs with this titanium oxide film. The results reveal that wettability, blood compatibility and fibroblasts compatibility of NiTi SMA are improved by the coating of titanium oxide film through H2O2 oxidation treatment. © 2007 The Nonferrous Metals Society of China.
    Original languageEnglish
    Pages (from-to)553-557
    JournalTransactions of Nonferrous Metals Society of China (English Edition)
    Volume17
    Issue number3
    DOIs
    Publication statusPublished - Jun 2007

    Research Keywords

    • biocompatibility
    • cytocompatibility
    • NiTi shape memory alloy
    • titanium oxide film
    • wettability