Hardware/Software Co-design of an ECG- PPG Preprocessor : A Qualitative & Quantitative Analysis

Research output: Chapters, Conference Papers, Creative and Literary WorksRGC 32 - Refereed conference paper (with host publication)peer-review

7 Scopus Citations
View graph of relations

Author(s)

Related Research Unit(s)

Detail(s)

Original languageEnglish
Title of host publication3rd International Conference on Electrical, Computer and Communication Engineering
Subtitle of host publicationECCE 2023
PublisherInstitute of Electrical and Electronics Engineers, Inc.
Number of pages6
ISBN (electronic)9798350345360
ISBN (print)979-8-3503-4537-7
Publication statusPublished - 2023

Publication series

NameInternational Conference on Electrical, Computer and Communication Engineering, ECCE

Conference

Title3rd International Conference on Electrical, Computer & Communication Engineering (ECCE 2023)
PlaceBangladesh
CityChittagong
Period23 - 25 February 2023

Abstract

This paper aims to design a digital system to pre-process electrocardiogram (ECG) and photoplethysmogram (PPG) signal for the purpose of hardware implementation. Muscle signal, motion artifacts, power line interference affect the biomedical signal during data acquisition. The proposed system focuses at removing the noises by designing infinite impulse response filter to remove power line noise and finite impulse response filter to eliminate other high and low frequency noises. At first the preprocessor is designed in Matlab to validate the simulation performance. Then the hardware is designed in xilinx system generator targeting Zedboard Zynq xc7z020-1clg484. Finally, we verified the hardware software codesign by comparing both outputs. For quantity based analysis different filtering techniques have been applied to determine the most optimized system in terms of resource utilization and power consumption. Pearson correlation coefficient of 0.9993 and 0.9982 have been found for ECG and PPG, respectively using Hamming filter technique for High and low pass filter. Root squared error for both signal has been also in the range of 10-2• These data validate the accuracy of the designed system providing quality assurance. Frequency spectrum also has been analyzed to ensure denoising of undesired signals. The designed preprocessor can be utilized for further analysis of the signals and designing digital systems & wearable devices for the detection of heart rate, cardiac diseases etc. © 2023 IEEE.

Research Area(s)

  • Electrocardiogram (ECG), Field Programmable Gate Array (FPGA), Photoplethysmo-gram (PPG)

Citation Format(s)

Hardware/Software Co-design of an ECG- PPG Preprocessor: A Qualitative & Quantitative Analysis. / Chowdhury, Aditta; Das, Diba; Cheung, Ray C.C. et al.
3rd International Conference on Electrical, Computer and Communication Engineering: ECCE 2023. Institute of Electrical and Electronics Engineers, Inc., 2023. (International Conference on Electrical, Computer and Communication Engineering, ECCE ).

Research output: Chapters, Conference Papers, Creative and Literary WorksRGC 32 - Refereed conference paper (with host publication)peer-review