Abstract
A two-stage Miller capacitance multiplier for compensation is proposed to implement an ultra-low frequency fully-integrated transimpedance amplifier (TIA). With the proposed multiplier, the TIA achieves a passband gain of 89dBΩ, a DC current rejection ranging from 0.5 to 85 μA, and a —3 dB highpass cutoff frequency of 0.5 to 9.5 Hz. This corresponds to an improvement of six times in DC current rejection range compared with previous techniques.
Citation Format(s)
Fully-integrated transimpedance amplifier for photoplethysmographic signal processing with two-stage Miller capacitance multiplier. / Li, Y.; Wong, A.K.Y.; Zhang, Y.-T.
In: Electronics Letters, Vol. 46, No. 11, 27.05.2010, p. 745-746.Research output: Journal Publications and Reviews (RGC: 21, 22, 62) › 21_Publication in refereed journal › peer-review