Skip to main navigation Skip to search Skip to main content

Frequency limitation in the calibration of microwave test fixtures

Research output: Journal Publications and ReviewsRGC 21 - Publication in refereed journalpeer-review

Abstract

The problem of frequency limitation arising from the calibration of asymmetric and symmetric test fixtures has been investigated. For asymmetric test fixtures, a new algorithm based on the thru-short-match (TSM) method is outlined. It is found that the conventional TSM method does not have any inherent frequency limitation, but using the same procedure with an unknown match may lead to the said problem. This limitation can be avoided by using a different algorithm. The various calibration methods for symmetric test fixtures using known standards are also discussed and the origin of the frequency limitation is identified. Several ways in avoiding the problem are proposed. There is good agreement between the theories and experimental data.
Original languageEnglish
Pages (from-to)2000-2006
JournalIEEE Transactions on Microwave Theory and Techniques
Volume51
Issue number9
DOIs
Publication statusPublished - Sept 2003

Research Keywords

  • Calibration
  • Deembedding
  • Microwave network analyzer
  • Scattering parameter measurement
  • Test fixture

Fingerprint

Dive into the research topics of 'Frequency limitation in the calibration of microwave test fixtures'. Together they form a unique fingerprint.