Fabrication of Multiple-Channel Electrochemical Microneedle Electrode Array via Separated Functionalization and Assembly Method

  • Xin-Shuo Huang
  • , Shuang Huang
  • , Shan-Tao Zheng
  • , Bao-Ming Liang
  • , Tao Zhang
  • , Wan Yue
  • , Fan-Mao Liu
  • , Peng Shi
  • , Xi Xie*
  • , Hui-Jiuan Chen*
  • *Corresponding author for this work

Research output: Journal Publications and ReviewsRGC 21 - Publication in refereed journalpeer-review

51 Downloads (CityUHK Scholars)

Abstract

Real-time monitoring of physiological indicators inside the body is pivotal for contemporary diagnostics and treatments. Implantable electrodes can not only track specific biomarkers but also facilitate therapeutic interventions. By modifying biometric components, implantable electrodes enable in situ metabolite detection in living tissues, notably beneficial in invasive glucose monitoring, which effectively alleviates the self-blood-glucose-managing burden for patients. However, the development of implantable electrochemical electrodes, especially multi-channel sensing devices, still faces challenges: (1) The complexity of direct preparation hinders functionalized or multi-parameter sensing on a small scale. (2) The fine structure of individual electrodes results in low spatial resolution for sensor functionalization. (3) There is limited conductivity due to simple device structures and weakly conductive electrode materials (such as silicon or polymers). To address these challenges, we developed multiple-channel electrochemical microneedle electrode arrays (MCEMEAs) via a separated functionalization and assembly process. Two-dimensional microneedle (2dMN)-based and one-dimensional microneedle (1dMN)-based electrodes were prepared by laser patterning, which were then modified as sensing electrodes by electrochemical deposition and glucose oxidase decoration to achieve separated functionalization and reduce mutual interference. The electrodes were then assembled into 2dMN- and 1dMN-based multi-channel electrochemical arrays (MCEAs), respectively, to avoid damaging functionalized coatings. In vitro and in vivo results demonstrated that the as-prepared MCEAs exhibit excellent transdermal capability, detection sensitivity, selectivity, and reproducibility, which was capable of real-time, in situ glucose concentration monitoring. © 2024 by the authors.
Original languageEnglish
Article number243
JournalBiosensors
Volume14
Issue number5
Online published13 May 2024
DOIs
Publication statusPublished - May 2024

Research Keywords

  • biosensing
  • blood glucose monitoring
  • micro/nanoneedle array
  • multi-channel electrochemical microneedle electrode array
  • separated functionalization and assembly

Publisher's Copyright Statement

  • This full text is made available under CC-BY 4.0. https://creativecommons.org/licenses/by/4.0/