Continuous Cuffless Blood Pressure Estimation Using Pulse Transit Time and Photoplethysmogram Intensity Ratio

Research output: Journal Publications and ReviewsRGC 21 - Publication in refereed journalpeer-review

257 Scopus Citations
View graph of relations

Author(s)

Detail(s)

Original languageEnglish
Article number7273864
Pages (from-to)964-972
Journal / PublicationIEEE Transactions on Biomedical Engineering
Volume63
Issue number5
Online published22 Sept 2015
Publication statusPublished - May 2016
Externally publishedYes

Abstract

Pulse transit time (PTT) has attracted much interest for cuffless blood pressure (BP) measurement. However, its limited accuracy is one of the main problems preventing its widespread acceptance. Arterial BP oscillates mainly at high frequency (HF) because of respiratory activity, and at low frequency (LF) because of vasomotor tone. Prior studies suggested that PTT can track BP variation in HF range, but was inadequate to follow the LF variation, which is probably the main reason for its unsatisfactory accuracy. This paper presents a new indicator, the photoplethysmogram intensity ratio (PIR), which can be affected by changes in the arterial diameter, and, thus, trace the LF variation of BP. Spectral analysis of BP, PTT, PIR, and respiratory signal confirmed that PTT was related to BP in HF at the respiratory frequency, while PIR was associated with BP in LF range. We, therefore, develop a novel BP estimation algorithm by using both PTT and PIR. The proposed algorithm was validated on 27 healthy subjects with continuous Finapres BP as reference. The results showed that the mean ± standard deviation (SD) for the estimated systolic, diastolic, and mean BP with the proposed method against reference were -0.37 ± 5.21, -0.08 ± 4.06, -0.18 ± 4.13 mmHg, and mean absolute difference (MAD) were 4.09, 3.18, 3.18 mmHg, respectively. Furthermore, the proposed method outperformed the two most cited PTT algorithms for about 2 mmHg in SD and MAD. These results demonstrated that the proposed BP model using PIR and PTT can estimate continuous BP with improved accuracy.

Research Area(s)

  • Arterial diameter change, cuffless blood pressure, photoplethysmogram intensity ratio, pulse transit time, respiration, vasomotion

Citation Format(s)

Continuous Cuffless Blood Pressure Estimation Using Pulse Transit Time and Photoplethysmogram Intensity Ratio. / Ding, Xiao-Rong; Zhang, Yuan-Ting; Liu, Jing et al.
In: IEEE Transactions on Biomedical Engineering, Vol. 63, No. 5, 7273864, 05.2016, p. 964-972.

Research output: Journal Publications and ReviewsRGC 21 - Publication in refereed journalpeer-review