Skip to main navigation Skip to search Skip to main content

An improved TM method for full two-port calibration of the asymmetric test fixtures

Research output: Journal Publications and ReviewsRGC 21 - Publication in refereed journalpeer-review

Abstract

Three known standards, including at least one transmission standard, are normally required for the full two-port calibration of test fixtures. Based on the triple-through method, a new general-purpose calibration procedure using only one known reflection standard is proposed in this paper. The experimental results show that our method can provide a simple and accurate approach to full two-port calibration of the asymmetric test fixtures. © 2005 Wiley Periodicals, Inc.
Original languageEnglish
Pages (from-to)438-441
JournalMicrowave and Optical Technology Letters
Volume45
Issue number5
DOIs
Publication statusPublished - 5 Jun 2005

Research Keywords

  • Calibration
  • Deembedding
  • Microwave network analyzer
  • Scattering parameter measurement
  • Test fixtures

Fingerprint

Dive into the research topics of 'An improved TM method for full two-port calibration of the asymmetric test fixtures'. Together they form a unique fingerprint.

Cite this