Skip to main navigation Skip to search Skip to main content

A simple and versatile 222Rn/220Rn exposure chamber

    Research output: Journal Publications and ReviewsRGC 21 - Publication in refereed journalpeer-review

    Abstract

    A simple 222Rn/220Rn exposure chamber is described. The chamber is made of acrylic and has a volume of 0.41 m3. The 222Rn and 220Rn sources can be operated independently or simultaneously. The concentrations of 222Rn and 220Rn were monitored by a continuous alpha spectrometer called RAD7. The capability of alpha spectrometry has removed the mutual interference between 220Rn and 222Rn, which is a major problem in measuring 220Rn. Furthermore, the capability can also provide fast response to changes in the 222Rn level and 220Rn level in the exposure chamber, with temporal resolutions of about 20 and 2min, respectively. © 2002 Elsevier Science B.V. All rights reserved.
    Original languageEnglish
    Pages (from-to)749-755
    JournalNuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
    Volume481
    Issue number1-3
    DOIs
    Publication statusPublished - 1 Apr 2002

    Research Keywords

    • CR39 detector
    • Detector sensitivity
    • LR115 detector
    • Radon
    • Thoron

    Fingerprint