A fast method to measure the thickness of removed layer from etching of SSNTD based on EDXRF

Research output: Journal Publications and ReviewsRGC 21 - Publication in refereed journalpeer-review

28 Scopus Citations
View graph of relations

Author(s)

Detail(s)

Original languageEnglish
Pages (from-to)161-164
Journal / PublicationRadiation Measurements
Volume36
Issue number1-6 SPEC.
Publication statusPublished - Jun 2003

Abstract

Solid state nuclear track detectors are commonly used for measurements of concentrations of radon gas and/or radon progeny. All these measurements depend critically on the thickness of the removed layer during etching. However, the thickness of removed layer calculated using the etching period does not necessarily provide a sufficiently accurate measure of the thickness. For example, the bulk etch rate depends on the strength of stirring during etching for the LR 115 detector. We propose here to measure the thickness of the removed layer by using energy-dispersive X-ray fluorescence spectrometry. In the present work, a reference silver nitrate pellet is placed beneath the LR 115 detector, and the fluorescence X-ray intensity for silver is then measured. We have found a linear relationship between the X-ray intensity and the thickness of the removed layer for LR 115 detector. This provides a fast method to measure the thickness of removed layer from etching of LR 115 detector. However, this method was found to be inapplicable for the CR-39 detector. Therefore, alternative methods have yet to be explored for the CR-39 detector. © 2003 Elsevier Ltd. All rights reserved.

Research Area(s)

  • Bulk etch, CR-39, EDXRF, LR 115, SSNTD

Citation Format(s)

A fast method to measure the thickness of removed layer from etching of SSNTD based on EDXRF. / Yip, C. W Y; Ho, J. P Y; Nikezic, D. et al.
In: Radiation Measurements, Vol. 36, No. 1-6 SPEC., 06.2003, p. 161-164.

Research output: Journal Publications and ReviewsRGC 21 - Publication in refereed journalpeer-review